Free mail service is available nationwide
Our services
Contact us
Our position
Links
Address: No. 137, south binzhou road, xiaodian district, taiyuan city
Zip code:030006
Telephone:400-0036-221
Question
Mail service
Answer
Question
Free technical support
Question
Free training
Taiyuan Taihang Pressure Test Technology Co., Ltd. 晋ICP备15007821号-1 Powered by 300.cn
Product description
The ADTS200C air data tester consists of a full-pressure sensor, a static pressure sensor, a display, a gas source, and an automatic pressure controller. It is mainly used to check the airtightness of aircraft instruments and connecting pipes after maintenance. It can also be used for testing and verification of airborne air data testers, altimeters, airspeed meters, Mach speedometers, dynamic pressure sensors, and static pressure sensors.
Characteristics
◆ Fully automatic control, just set the target value
◆ Independent Ps, Pt dual channel automatic control design
◆Providing a hand-held device, the user can control the instrument from a distance
◆Using portable luggage chassis installation method, convenient for users to use in the field
◆Good man-machine interface, panel setting parameters
◆Pressure unit and aviation unit display user-friendly readings under the same interface
◆The gas source provides pressure source and vacuum gas source
◆ Provide external vacuum source interface, pressure source interface
◆The test system gas path adopts Φ7 pagoda interface, which is convenient to connect with the tested table.
◆Using RS232 standard communication interface for data transmission
Technical indicators
Measurement parameter
Measuring range
Resolution
Maximum allowable error
Static pressure (hPa)
25~1200
0.01
0.02 level
Full pressure (hPa)
25 to 3000
0.01
0.02 level
Height (m)
-1000~25000
0.1
-1000
±1.8
0
±2
9000
±5.3
18000
±20.3
25000
±61.7
Indicating airspeed (km/h)
50~1600
0.1
50
±9.6
800
±0.5
1600
±0.2
Mach number M
0.1~5.0000
0.0001
±(0.0001~0.005)
Climb rate (m/s)
±300
0.1
±1%FS or ±1m/s (large value)
Rate of change of speed
(km/h/s)
0~21
0.1
±1%FS
Control stability
±0.003%
Environmental requirements
◆Working temperature: -30 ° C ~ 55 ° C
◆Storage temperature: -55 ° C ~ 70 ° C
◆Working humidity: no more than 95% RH
◆Dimensions: 580 mm × 460 mm × 270 mm
◆Power supply: AC 220V±22V; 1.5A or: DC28V (18V~36V); 9A
◆Weight: no more than 25kg
ADTS300 Atmospheric Data Tester
Next article